Overview
The Edwards Diffstak 100 (ISO100 inlet) design has been proven over many years with thousands of pumps installed.
This pump has a manually operated high vacuum isolation valve and is supplied with: NW25 elbow, NW25 centering-ring, NW25 clamp, water pipe couplings and ferrules, inlet ISO 100 Co-Seal.
Technical Data
| Pumping speed | |
| nitrogen | 280 l s-1 |
| hydrogen | 500 l s-1 |
| Min backing pump displacement* | 5 m3h-1 |
| Recommended backing pump | RV5 or RV8 |
| Recommended fluid | Santovac® 5 |
| Fluid charge (dry) | 125 ml |
| Inlet connection compatible with | ISO100 |
| Backing connection | NW25 |
| Cooling water connection | 6 mm compression fittings |
| Heater power | 0.65 kW |
| Min cooling-water flow at 20°C | 60 l h-1 |
| Pneumatic connections† | 6mm coupling x ¼ BSP male stud |
| Pneumatic actuating pressure | |
| Minimum | 2.4bar / 35psi |
| Maximum | 6.9bar / 100psi |
| Weight | |
| 100/300M | 12kg |
| 100/300P | 13kg |
* For maximum throughput
† 100/300P only
Inquiry - Diffstak 100/300M, 110-125V, 1-ph, 50/60Hz